RBS3202 Test with CMU300 and RNC/Iub SIM
1CM56_E 3 Rohde & Schwarz
2 Operating Principle
RF Tester
R&S CMU300
Node B
Ericsson RBS3202
NetHawk
RNC/Iub Simulator
Iub
Uu
Before the R&S CMU300 can test the RF parameters according to [1], the
Node B must be altered to a test mode with the above-mentioned test
scenarios. In most cases these test modes are not supported by the radio
network controller (RNC) of the operative network. In this particular case
the NH RNC/Iub Simulator assumes this role. The BTS, being isolated
from the network, is activated by the NH RNC/Iub Simulator via the Iub
interface. The BTS is then tested as follows:
Transmission parameters:
• The NH RNC/Iub Simulator activates an appropriate test model on the
downlink (DL)
• The R&S CMU300 measures the transmission parameters (RF power,
modulation, spectrum, code domain measurements)
Reception parameters:
• The NH RNC/Iub Simulator configures the BTS receiver to receive a
RMC
• The BTS provides the TTI trigger signal
• The CMU300 transmits the requested PN modulated RMC; the start of
transmission is determined by the TTI signal
• The BER is evaluated (transport level) in the NH RNC/Iub Simulator