RBS3202 Test with CMU300 and RNC/Iub SIM
1CM56_E 9 Rohde & Schwarz
• softkey Analyzer Settings: Set the DL frequency or channel
number to be used on the RNC simulator.
• softkey Exp. Pow. Trigger: Switch to manual control (mode:
Manual) and set the expected power approx. 13 dB (PEP) higher
than the maximum Node B output power set on the RNC. If the
message "Overload at connector RF1" occurs, increase the value
of the expected input power until the message ceases to occur.
• Carry out power measurement.
• Switch to modulation analysis (softkeys Menus / Modulation) and set
the activated DL test model (softkeys Analyzer Settings / Test Model).